General Technical Data
Weight (oz): (with probe) Approx. 25 / (with probe) Approx. 700 g
Dimensions (in): 8 x 3 x 2 / 201 x 83 x 44 mm
Operating Temperature: 23° to 113 °F / -5 to +45 °C
Display Type: LCD
Display Function: Backlit 2-line display
Power Supply: Battery: 1500 mAh, mains unit 5V/1A
Storage Temperature: -4° to 122 °F / -20 to +50 °C
Temperature - Type K TC
Measuring Range: -4° to 212 °F / -20 to +100 °C
Accuracy: ±2 °F / ±1 °C
Resolution: 0.1 °F / 0.1 °C
Reaction Time: < 50 sec
Temperature (ambient temperature)
Temperature - Type J TC
Measuring Range: 32° to 752 °F / 0 to +400 °C
Accuracy: ±1.8 °F (32° to 212 °F) / ±1 °C (0 to +100 °C) / ±1.5 % of mv (> 212 °F) / ±1.5 % of mv (> 100 °C)
Resolution: 0.1 °F / 0.1 °C
Reaction Time: < 50 sec
Temperature (flue gas): Flue gas O₂
Measuring Range: 0 to 21 Vol.%
Accuracy: ±0.2 Vol.%
Resolution: 0.1 Vol.%
Reaction time t₉₀: 30 sec
Flue Gas Draft
Measuring Range: -8 to +8 InH₂O / -20 to +20 hPa
Accuracy: ±0.01 InH₂O (-1.20 to +1.20 InH₂O) / ±0.03 hPa (-3.00 to +3.00 hPa) / ±1.5 % of mv (Remaining Range)
Resolution: 0.01 InH₂O / 0.01 hPa
Efficiency (calculated)
Measuring Range: 0 to 120 %
Resolution: 0.1 %
Flue Gas loss (calculated)
Measuring Range: 0 to 99.9 %
Resolution: 0.1 %
Flue Gas CO₂ calculation (calculated from O₂)
Measuring Range: 0 to CO₂ max (Display range)
Accuracy: ±0.2 Vol.%
Resolution: 0.1 Vol.%
Reaction time t₉₀: < 40 sec
Pressure measurement
Measuring Range: -16 to +16 InH₂O / -40 to +40 hPa
Accuracy: ±0.2 InH₂O / ±0.5 hPa
Resolution: 0.1 InH₂O / 0.1 hPa
Flue gas CO (without H₂-compensation)
Measuring Range: 0 to 4000 ppm
Accuracy: ±20 ppm (0 to 400 ppm) / ±5 % of mv (401 to 2000 ppm) / ±10 % of mv (2001 to 4000 ppm)
Resolution: 1 ppm
Reaction time t₉₀: 60 sec
Ambient CO
Measuring Range: 0 to 4000 ppm
Accuracy: ±20 ppm (0 to 400 ppm / ±5 % of mv (401 to 2000 ppm) / ±10 % of mv (2001 to 4000 ppm)
Resolution: 1 ppm
Reaction Time: 60 sec